Operation in Measurement Mode
29
Operating Instructions Vamos, 1. US ed.
Selecting parameters to measure
Calibration
Selecting parameters to measure
Monitoring and presentation of anesthetic agents, N2O, CO2
values and SpO2, Pulse values can be switched on and off.
The default for both settings on power-up is »on«.
To set:
Turn or press dial knob to select main menu.
Turn dial knob to select »Param.« parameter.
Press dial knob to open the parameter menu.
Turn dial knob to select setting for gas measurement
parameters.
Press dial knob to confirm setting.
The setting is marked with an arrow.
Turn dial knob to select setting for SpO2 and Pulse
measurement.
Press dial knob to confirm setting.
The setting is marked with an arrow.
Press dial knob to return to the main menu.
Press dial knob again to return to the measurement
display.
If a parameter is switched off, the respective values are
replaced by "XX". If the gases measurement is switched off,
the field for the CO
2-curve shows "XX" in its center.
Calibration
SpO2 measurement does not need to be calibrated.
Gas measurement is calibrated every 2 hours by the monitor.
During calibration, which lasts approximately 30 seconds, the
reading »CAL« appears in place of the gas measurement
values.
Gas measurement calibration may also be started manually
from the main menu:
Turn or press dial knob to select main menu.
Turn dial knob to select »Cal« in the main menu.
Press dial knob to confirm.
The device calibrates automatically; the next automatic
calibration is carried out again after 2 hours.
onGases
off
Fi
Fet
Hal
[%]
N2O
[%]
CO2
[%]
[1/min]
[%]
on
SpO2
Pulse
98
110
Param.
0.6
0.1
30
25
0.0
4.5
off
SpO2
exit
RETURN TO THIS MANUAL'S TABLE OF CONTENTS