Dräger X-dock
Frequently Asked Questions
HOME
Tubes must be flooded so that the test gas reaches the module. Optimized and
the shortest possible paths are used in X-dock to keep gas consumption low.
However, long external supply lines require increased flooding to ensure the
device is ready for testing quickly. This option is available to make sure this
happens only with long supply lines and gas is not wasted.
M
ISCELLANEOUS
What is the purpose of the option “more than 2m hose length”?
Since it is not possible to charge X-am 125 devices while they are being used,
the charging process will not start immediately after closing the lid, but after
a delay of 15 minutes. The counter starts again if a test is carried out within
these 15 minutes. The charging circuit turns on and the device turns off only
after 15 minutes without interaction with the corresponding module. The lid must
be closed for this!
This type of operation is recommended especially with group and scheduler
mode.
When does charging start at the X-am 125+ module?
9094227_X-dock_FAQs_210x297_ENGL_5 20.11.13 12:11 Seite 14

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-dock and is the answer not in the manual?

Dräger X-dock Specifications

General IconGeneral
Power Supply100 – 240 VAC, 50 – 60 Hz
Humidity10 to 95 % RH, non-condensing
Data StorageInternal memory for device data, calibration data, and test results
InterfaceUSB, Ethernet
Device TypeAutomated test and calibration station
CalibrationAutomatic calibration
Bump TestAutomatic bump test
Supported GasesDepends on the gas detector being tested (e.g., H2S, CO, O2, LEL)
CompatibilityCompatible with Dräger portable gas detectors