Dräger X-dock
Frequently Asked Questions
MISCELLANEOUS Can I use modules with different firmware at one station?
X-dock requires modules and masters with the same firmware. It is not possible
to use modules and masters within a station that do not use the same firmware;
this results in an error message. In this case, please perform a simple USB
update. Modules that have not been recognised prior are now updated to the
same firmware version.
However, it is possible to combine different firmware versions of different
stations in a network. Nevertheless, this is not recommended and the latest
supported firmware should be used on all stations.
HOME
Which type of printers is supported by the X-dock station?
The X-dock Manager certainly supports all printers that are triggered via the
Windows operating system. If you want to print directly with the Dräger X-dock
station please ensure that the printer connected via an USB stick is “postscript
compliant”. Please contact your Dräger representative for a complete list of
supported printers.
The quick bump test checks exceeding of the first alarm level. As soon as this
limit was exceeded the test is regarded as passed. In many cases this test is
sufficient.
The extended bump test goes one step further – it checks the accuracy of the
sensors every time. This means it is checked whether the value appears in
a window around the concentration of the test gas.
What is the difference between the quick and the extended bump test?
9094227_X-dock_FAQs_210x297_ENGL_5 20.11.13 12:10 Seite 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger X-dock and is the answer not in the manual?

Dräger X-dock Specifications

General IconGeneral
Power Supply100 – 240 VAC, 50 – 60 Hz
Humidity10 to 95 % RH, non-condensing
Data StorageInternal memory for device data, calibration data, and test results
InterfaceUSB, Ethernet
Device TypeAutomated test and calibration station
CalibrationAutomatic calibration
Bump TestAutomatic bump test