Continuous Flow Rate Setting cf =
2 LPM +/- 0.4 LPM
Deviation of delivered oxygen per breath =
+/- (1 ml/breath + 10% of the set initial value
per breath)
Tested per ISO 80601-2-67:2014 Sec 201.12.1.101
Tolerance on measurements = +/- 15%
(including measurement uncertainty)
SPECIFICATIONS
All volumes and flows are expressed at STPD
(standard temperature and pressure, dry)