The following are required for disinfection/clean-
ing:
ü
Non-foaming disinfectant/cleaning agent that
is compatible with the materials.
ü
Unit care system, e.g. OroCup
To pre-clean, suck up 2 litres of water with the
care system.
Aspirate the disinfection/cleaning agent with
the care system.
13.3 Once or twice a week before
the midday break
Under harsher conditions (e.g. hard water
or frequent use of prophy powders) 1x
daily before the midday break
The following are required for cleaning:
ü
Special non-foaming detergent for suction
units that is compatible with the materials.
ü
Unit care system, e.g. OroCup
To pre-clean, suck up 2 litres of water with the
care system.
Aspirate the cleaning agent with the care sys-
tem.
Rinse with ca. 2 l water after the application
time.
14 Replace the amalgam col-
lector vessel
NOTICE
Risk of contamination if the amalgam
collector vessel is reused since the
collector vessel is not water-tight.
Do not use the collecting container
more than once (disposable item).
Wear protective equipment to avoid any
risk of infection (e.g. liquid-tight protective
gloves, protective goggles, face mask).
We strongly recommend that the amal-
gam collecting container should only be
changed in the morning before the start
of work. This will prevent fluid from drip-
ping out of the drum while it is being
changed.
Disconnect all power from the unit.
Remove the full amalgam collecting container
and from the device.
Pour disinfectant for suction units (e. g. Orotol
plus, 30 ml) into the full amalgam collecting
container.
Close and secure the full amalgam collecting
container using the cap. Observe the markings
on the cap and on the collecting container.
Place the securely closed amalgam collecting
container into its original packaging and seal.
Insert a new amalgam collecting container in
the unit and clamp it in position.
Only use original amalgam collecting con-
tainer.
Switch on the power supply. The unit is ready
for operation again.
14.1 Disposal of amalgam collect-
ing container
The contents of the amalgam collecting
container are contaminated with heavy
metals and must not be disposed of as
household waste or the environment.
Collection and waste disposal by a waste
management company specialised in surgery
waste.
Collection and waste disposal by an approved
waste management company.
Usage
9000-606-31/30 2101V005 31
EN

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental VSA 300 S and is the answer not in the manual?

Durr Dental VSA 300 S Specifications

General IconGeneral
CategoryDental equipment
Number of treatment units1
Voltage230 V
Frequency50/60 Hz
Power consumption1200 W
Weight45 kg
Max. flow rate300 l/min