Operating Manual
English
R

Questions and Answers

  • J
    Justin KnightAug 10, 2025
    Why does the CAL symbol appear on my Dynatron AG Test Equipment?
    • D
      Dominique PageAug 10, 2025
      If the CAL symbol appears on your Dynatron AG Test Equipment OxySpy, it could be due to a few reasons. It might be because a new battery or a new O sensor has been installed. Another reason could be that the calibration was not successful. In any of these cases, you should calibrate the OxySpy.
  • K
    kathleen00Aug 25, 2025
    What to do if my Dynatron AG Test Equipment OxySpy does not see an O sensor?
    • J
      Jimmy DavisAug 25, 2025
      If your Dynatron AG Test Equipment OxySpy doesn't detect an O sensor, there are several possible causes. First, ensure that an O sensor is properly installed and plugged in. Another reason could be water in the housing. In this case, remove the battery and the O sensor, then dry the case with a towel and warm air (at a maximum of 50 °C / 122 °F). If the problem persists, there may be a plug or wire defect, or an electronic defect. In these cases, take the OxySpy to the store that it was purchased from.

Dynatron AG OxySpy Specifications

General IconGeneral
TypeOxygen Analyzer
Measurement Range0-100% O2
Accuracy±1% of full scale
Resolution0.1% O2
Response Time<15 seconds
Operating Temperature0°C to 50°C
Power Supply9V battery
Weight250 g