Maintenance and Storage
20
• Protect the unit from impact and intensive sun
and heat.
• Store the unit in a cool, not too dry place. This
helps to preserve the O
2
sensor life.
• Ideal storage temperature: 5 °C up to +15 °C
(41 °F up to 59 °F)
• Do not leave the OxySpy in an oxygen enriched
atmosphere except when it is necessary.
• Disconnecting the
O
2
Sensor from the socket
does not extent its life.
• Replace the battery and
O
2
sensor early.
• Do not clean the unit with any solvent (except
water) or detergent.
• Clean the OxySpy with a damp, soft towel.
• Dry the instrument with a soft towel.
• Do not scratch the display window.
• Maximum altitude: 4000 m (13,000 feet)
above sea level
• Temperature range: 0 °C up to +50 °C
(32 °F to 122 °F)
• Relative air humidity: 10-90%,
non-condensing
•O
2
Display: • Resolution: 0.1% O
2
• Display range: 0.0 up to 199.9%O
2
•O
2
Sensor: • Type Dynatron P/N DiveceL3.D1
or other O
2
sensors if licensed by
the manufacturer.
• Lifetime: minimum 1 year*
• Battery: • Type: CR2430
• Lifetime: min. 3 years at
500 on/ off cycles per year
• Linearity: <3% (at constant pressure and
temperature)
Maintenance
and Storage
English
Operational
limits
Technical
Information
*
Lifetime depending on environment conditions and the used product.