Set Up Procedure / Maintenance
8
1. Open the rear side of the housing according
to the chapter "Opening the OxySpy" (page 7).
Consider the safety considerations on page 5
and the ones on the
O
2
sensor and its package.
2. Take the
O
2
sensor (if there is already one) out
of its position and pull off the mounting ring.
If the
O
2
sensor is disconnected while the unit is
on, the display shows the word "no" and the
O
2
sensor symbol.
In this situation the OxySpy switches off after 5
minutes automatically.
3. Carefully pull the plug out of the socket. The
tab, which stops the plug from being disconnec-
ted, has to be pulled back while pulling out the
plug.
Do not pull on the wires!
Do not open the airtight container before you
need to use the
O
2
sensor.
After opening the sealed package the
O
2
sen-
sor needs 30 minutes start-up time. This
means that calibration cannot be made until 30
minutes after opening the package. See "Cali-
bration" page 11.
Take care of the environment; properly dispose
of used
O
2
sensor! The sensor should be dis-
posed of according to local laws.
Fitting the
O
2
Sensor
English
O
2
sensor
Mounting ring