Hangzhou Zhongchuang Electronics Co., Ltd.
23
Capacitance C and loss D
100Hz/120Hz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
20mF
4.000mF∼ 20.000mF
5.00%+5 digits
0.0500
Series
4mF
400.0μF∼ 3.9999mF
1.00%+3 digits
0.0100
Series
400μF
40.00μF∼ 399.99μF
0.30%+2 digits
0.0030
Series
40μF
4.000μF∼ 39.999μF
0.20%+2 digits
0.0020
Series
4μF
400.0nF∼ 3.9999μF
0.20%+2 digits
0.0020
----
400nF
40.00nF∼ 399.99nF
0.2%+2 digits
0.0020
Parallel
40nF
4.000nF∼ 39.999nF
0.3%+3 digits
0.0030
Parallel
4nF
0pF∼ 3.999nF
1.2%+5 digits
------
Parallel
1kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
1000uF
400.0uF∼ 999.99uF
2.00%+5 digits
0.0200
Series
400μF
40.00μF∼ 399.99μF
1.00%+3 digits
0.0100
Series
40μF
4.000μF∼ 39.999μF
0.30%+2 digits
0.0030
Series
4μF
400.0nF∼ 3.9999μF
0.20%+2 digits
0.0020
----
400nF
40.00nF∼ 399.99nF
0.2%+2 digits
0.0020
Parallel
40nF
4.000nF∼ 39.999nF
0.2%+3 digits
0.0030
Parallel
4nF
400.0pF∼ 3.9999nF
0.3%+3 digits
0.0030
Parallel
400pF
0.0pF~399.9pF
1.2%+5 digits
------
Parallel
10kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
100μF
40.00μF∼ 100.00μF
3.00%+5 digits
0.0300
Series
40μF
4.000μF∼ 39.999μF
1.0%+3 digits
0.0100
Series
4μF
400.0nF∼ 3.9999μF
0.30%+2 digits
0.0030
Series
400nF
40.00nF∼ 399.99nF
0.2%+2 digits
0.0020
Series
40nF
4.000nF∼ 39.999nF
0.2%+2 digits
0.0020
------
4nF
400.0pF∼ 3.9999nF
0.2%+2 digits
0.0020
Parallel
400pF
40.00pF~399.99pF
0.3%+3 digits
0.0030
Parallel
40pF
0.00pF~39.99pF
1.2%+5 digits
------
Parallel
40kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended