EasyManua.ls Logo

East Tester ET430 - Page 27

East Tester ET430
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hangzhou Zhongchuang Electronics Co., Ltd.
23
Capacitance C and loss D
100Hz/120Hz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
20mF
4.000mF 20.000mF
5.00%+5 digits
0.0500
Series
4mF
400.0μF 3.9999mF
1.00%+3 digits
0.0100
Series
400μF
40.00μF 399.99μF
0.30%+2 digits
0.0030
Series
40μF
4.000μF 39.999μF
0.20%+2 digits
0.0020
Series
4μF
400.0nF 3.9999μF
0.20%+2 digits
0.0020
----
400nF
40.00nF 399.99nF
0.2%+2 digits
0.0020
Parallel
40nF
4.000nF 39.999nF
0.3%+3 digits
0.0030
Parallel
4nF
0pF 3.999nF
1.2%+5 digits
------
Parallel
1kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
1000uF
400.0uF 999.99uF
2.00%+5 digits
0.0200
Series
400μF
40.00μF 399.99μF
1.00%+3 digits
0.0100
Series
40μF
4.000μF 39.999μF
0.30%+2 digits
0.0030
Series
4μF
400.0nF 3.9999μF
0.20%+2 digits
0.0020
----
400nF
40.00nF 399.99nF
0.2%+2 digits
0.0020
Parallel
40nF
4.000nF 39.999nF
0.2%+3 digits
0.0030
Parallel
4nF
400.0pF 3.9999nF
0.3%+3 digits
0.0030
Parallel
400pF
0.0pF~399.9pF
1.2%+5 digits
------
Parallel
10kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended
100μF
40.00μF 100.00μF
3.00%+5 digits
0.0300
Series
40μF
4.000μF 39.999μF
1.0%+3 digits
0.0100
Series
4μF
400.0nF 3.9999μF
0.30%+2 digits
0.0030
Series
400nF
40.00nF 399.99nF
0.2%+2 digits
0.0020
Series
40nF
4.000nF 39.999nF
0.2%+2 digits
0.0020
------
4nF
400.0pF 3.9999nF
0.2%+2 digits
0.0020
Parallel
400pF
40.00pF~399.99pF
0.3%+3 digits
0.0030
Parallel
40pF
0.00pF~39.99pF
1.2%+5 digits
------
Parallel
40kHz
Range
Range of display
Accuracy Ce
Accuracy De
Equivalent mode
recommended

Related product manuals