EasyManua.ls Logo

Easymax MU - Understanding Your Test Results; Unusual Test Results

Easymax MU
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
Understanding Your Test Results
The EASYMAX
®
MU Blood Glucose test strips are whole-blood referenced and calibrated for easier
comparison to lab results. The American Diabetes Association recommends a post-meal glucose level of less
than 180 mg/dL (10 mmol/L) and a pre-meal glucose of 70–130 mg/dL (3.88-7.22 mmol/L)*.
Reference: American Diabetes Association. Standards of medical care in diabetes-Table 10. Diabetes
care. 2011; Vol. 34, Suppl. 1, S21.
Unusual Test Results
If your test result does not match the way you feel, please follow these steps:
1. Run a control test, Chapter 2, "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3, "Testing Your Blood glucose."
3. If your test results still do not reflect the way you feel, call your doctor immediately.
1. Extremely high humidity may affect the test results. A relative humidity greater than 90%
may cause inaccurate results.
2. Hematocrit below 20% may cause higher results. Hematocrit above 60% may cause
lower results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test
near an operating microwave oven.

Related product manuals