49
3. Critically ill patients should not be tested with EASYMAX
®
MU glucose meters.
4. DO NOT use during xylose absorption testing. Xylose in the blood will interfere Self-Monitoring Blood
Glucose System.
5. Interfering Substances depend on the concentration. The below substances up to the test concentration
will not affect the test results.
80 mg/dL
(4.4 mmol/L)
250 mg/dL
(13.9 mmol/L)
500 mg/dL
(27.8 mmol/L)
Ascorbic Acid, 4 mg/dL (0.26mmol/L) 10.89% -1.76% 4.55%
Ibuprofen, 50 mg/dL (2.43mmol/L) 3.10% 2.88% 4.62%
L-Dopa, 1.8 mg/dL (0.09mmol/L) 10.59% 7.91% 4.90%
Sodium Salicylate, 50 mg/dL (3.12mmol/L) -2.59% 9.42% -0.84%
Tetracycline, 1.5 mg/dL (0.03mmol/L) -5.32% 3.81% 3.20%
Tolbutamide, 100 mg/dL (3.7mmol/L) -2.60% 12.30% 0.89%
Bilirubin-unconjugated, 2.4 mg/dL (0.04mmol/L) -2.52% 4.05% -0.23%
Uric acid, 8 mg/dL (0.48mmol/L) 2.71% 9.55% -1.75%
Xylose, 4 mg/dL (0.27mmol/L) -5.12% -1.64% -4.44%
Glucose
Level
Bias
Concentrations
of the interference tested