Cleaning Instruction: All blood and other body fluids must be thoroughly cleaned from surfaces and
objects before disinfection by the germicidal wipe. Open, unfold and use first germicidal wipe to remove heavy
soil after every use.
Disinfection Instruction:
Unfold a clean wipe and thoroughly wet all the surface of the meter, including the
strip port and other connection port. Unfold a clean wipe and thoroughly wet all the surface of the lancing
device, including cap or AST cap if used. Treated area must remain visibly wet for a full 2 minutes. Use
additional wipe(s) if needed to assure continuous 2 minute wet contact time. Let the devices air dry for 30
seconds. Disinfect at least once a week.
Do Not:
Get any moisture in the test strip slot.
Spray any cleaning solution directly onto the
meter.
Put the meter under water or liquid.
Pour liquid into the meter.
Do:
Make sure the meter is turned off during cleaning and
disinfection.
Keep the test strip vial tightly closed when performing
the cleaning and disinfection procedures because the
fumes from the disinfectant may affect the
performance of the strip.
After cleaning or disinfection, please perform the
physical appearance and performance check of
devices.
Consult cleaning instruction of single use damp cloth.