Cleaning And Disinfection Frequency
At least 14,600 validated cycles for the meter and lancing device were done and 10,000 cycles of cleaning
and 4,600 cycles of disinfections were included. 4 cycles of cleaning per day and 1 cycle of disinfection per
week allow the user to use the device for 4 years.
Cleaning Frequency : 4 cycles/day X 365 days/year X 4 year warranty = 5840 cycles of cleaning
<10,000 validated cycles of cleaning.
Disinfection Frequency: 1 cycle/week X 52 weeks/year X 4 year warranty = 208 cycles of cleaning
< 4,600 validated cycles of disinfection
4 year warranty for properly cleaning and disinfection. After 4 years, the meter must be replaced with a
new meter.