E C O P H Y S I C S Operating Guide
CLD 780 TR / July 2000 55
(26) “Stand-by Mode“-window
The analyzer indicates that it is in a “Stand-by“ state.
Exit “Stand-by“ mode with “MEAS“; after a short “O3UP“ phase the ana-
lyzer is ready to resume measurement.
(27) “Stand-by Mode DOWN“-window
“Stand-by“ state with an error or warning code after a serious problem
has occurred.
After a serious problem has occurred (E-01 to E-05, see 10.1.1) the
measuring function is automatically interrupted. The analyzer is put in the
“Stand-by“ mode and “DOWN“ appears in the upper right hand corner
of the display. By pushing the “SYS“ key the error message text is shown
in the display. After rectifying the error and pushing the “MEAS“ key the
message will disappear and measurement will resume. The short “O3UP“
and “PWUP“ phases precede the normal measuring mode “MEAS“.
Return with “MEAS“.