CHAPTER 4, DIGITAL COMMUNICATION
98307600 Rev. C-1 4-19
Format
DEVENT,{<DEVICE I.D.>}{TERMINATOR}
Device response
#XX {Message #XX <CR><LF> OCCURRED AT HH:MM DD-MON-YY}<CR><LF>
where:
The last 100 messages are reported. XX is the index into the event log message
buffer; 99 = oldest point (reported first), 0 = newest point (reported last).
The message field is null if no message exists.
Command
DGAIN
Function
Dumps instrument gain data to the serial port.
Format
DGAIN,{<DEVICE I.D.>}{TERMINATOR}
Device response
{INSTRUMENT GAIN}<CR><LF>
All numbers are in floating point format.
Command
DINSTR
Function
Dumps the INSTRUMENT STATUS MENU variables to the serial port.
Format
DSTATUS,{<DEVICE I.D.>}{TERMINATOR}
Device response
{STAT1},{STAT2},{STAT3},{STAT4},{STAT5},{STAT6},{STAT7},{STAT8},
{STAT9},{STAT10},{STAT11},{STAT12}<CR><LF> where:
{STAT1} = gas flow
{STAT2} = gas pressure
{STAT3} = reference voltage
{STAT4} = concentration voltage

Table of Contents