EasyManuals Logo

Ecotech A Series User Manual

Default Icon
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
EC9830 CO ANALYZER OPERATION MANUAL
4-20 98307600 Rev. C-1
{STAT5} = analog supply
{STAT6} = digital supply
{STAT7} = ground offset
{STAT8} = ozone generator flow
{STAT9} = high voltage
{STAT10} = lamp current, mA
{STAT11} = ambient pressure
{STAT12} = Startup flag (1 = in startup mode).
Startup flag and ground offset are integers; all other numbers are in floating point
format. The status field is null if it does not apply to the analyzer type.
Command
DSPAN
Function
Commands the unit under test to enter the span mode and stay there.
Format
DSPAN,{<DEVICE I.D.>}{TERMINATOR}
Device response
<ACK> if the unit under test is able to perform the command, <NAK> if not.
Command
DTEMPS
Function
Dumps the SYSTEM TEMPERATURES MENU variables to the serial port.
Format
DTEMPS, [,DEVICE I.D.>}{TERMINATOR}
Device response
{TEMP1},{TEMP2},{TEMP3},{TEMP4},{TEMP5},{TEMP6},{TEMP7},{TEMP8},
{TEMP9},{TEMP10}<CR><LF> where:
TEMP1 = cell temperature
TEMP2 = converter temperature
TEMP3 = chassis temperature
TEMP4 = flow temperature
TEMP5 = cooler temperature
TEMP6 = mirror temperature
TEMP7 = lamp temperature

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecotech A Series and is the answer not in the manual?

Ecotech A Series Specifications

General IconGeneral
BrandEcotech
ModelA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals