EC9830 CO ANALYZER OPERATION MANUAL 
2-6  98307600 Rev. C-1 
Desired Output (Volts)  Terminating Resistance (Ohms) 
10 Volts  500 Ohms 
5 Volts  250 Ohms 
1 Volt  50 Ohms 
0.1 Volt  5 Ohms 
When using voltage output, the source resistance is 1000 ohms. The recorder or 
DAS input resistance should be greater than 500K ohms for a measurement error 
no greater than 1%. 
 
2.1.2.2 Sample Gas Connections 
Caution 
Sample  connections  to  the  EC9830  should  be 
maintained at ambient pressure, with any excess 
flow vented to the atmosphere. 
The EC9830 requires at least 1.50 slpm (1.0 slpm sample plus 50% overflow) of 
particulate-filtered  (<5 micron),  dry  (noncondensing)  sample  furnished  at  all 
times. A 5 micron inlet filter is necessary to meet USEPA requirements which is 
already installed in the A series analyzer. 
Tubing  used  for  sample  gas  and  exhaust  connections  must  be  ¼ inch  OD  and 
1/8 to  3/16 inch  ID.  The  recommended  ID  is  5/32 inch.  A  segment  of  clean 
Teflon® tubing should be purchased to connect the sample source to the sample 
inlet. Only use lines and fittings made of stainless steel, Teflon, Kynar®, or glass. 
Instructions for Kynar fittings: 
  Cut the tubing squarely and remove any burrs. 
  Insert the tubing through the back of the nut until it reaches the tube stop in 
the fitting. 
  Tighten the nut finger-tight plus 1-1/2 to 2 turns. A squeaking sound when 
tightening the nut is normal. 
  All  nuts  should  be  re-tightened  when  the  system  reaches  operating 
temperature.