CHAPTER 3, CALIBRATION 
98307600 Rev. C-1  3-5 
3.2.1.2.5 Output Manifold 
The output manifold should be of sufficient diameter to ensure an insignificant 
pressure drop at the analyzer connection. The system must have a vent designed to 
ensure  atmospheric  pressure  at  the  manifold  and  enough  inlet  flow  to  prevent 
ambient air from entering the manifold. 
3.2.1.3 Reagents 
3.2.1.3.1 CO Concentration Standard(s) 
Cylinder(s) of CO in air containing appropriate concentration(s) of CO suitable for 
the selected operating range of the analyzer under calibration; CO standards for the 
dilution method may be contained in a nitrogen matrix if the zero air dilution ratio 
is  not  less  than  100:1. The  assay of the  cylinder(s) must  be traceable either to  a 
National  Institute  of  Standards  and  Technology  (NIST)  CO  in  air  Standard 
Reference  Material  (SRM)  or  to  an  NIST/EPA  approved  commercially  available 
Certified  Reference  Material  (CRM).  CRM’s  are  described  in  Calibration 
Reference 2,  and  a  list  of  CRM  sources is  available  from  the  address  shown for 
Calibration Reference 2. A recommended protocol for certifying CO gas cylinders 
against either a CO SRM or a CRM is given in Calibration Reference 1. CO gas 
cylinders should be recertified on a regular basis as determined by the local quality 
control program. 
3.2.1.3.2 Dilution Gas (Zero Air) 
Zero  air  is  described  as  air  free  of  contaminants  which  will  cause  a  detectable 
response  on  the  CO  analyzer.  The  zero  air  should  contain  <0.1 ppm  CO.  A 
procedure  for  generating  CO  zero  air  is  given  in  Calibration  Reference 1;  the 
Ecotech GasCal 1000 calibrator and 8301 Zero Air source calibrator has CO-free air 
capability sufficient for this operation. 
3.2.1.4 Procedure Using Dynamic Dilution Gas Method 
1. Assemble a dynamic calibration system such as the one shown in Figure 3-1. 
All  calibration  gases including zero air must be introduced into the sample 
inlet of the analyzer system. The EC9830 instrument sample inlet is labeled 
Inlet and is located on the rear panel. 
2. Ensure that all flowmeters are properly calibrated, under the conditions of use, 
if appropriate, against an authoritative standard such as a soap-bubble meter 
or wet-test meter. All volumetric flow rates should be corrected to 25° C and 
760 mm Hg (101 kPa). A discussion on calibration of flowmeters is given in 
Calibration Reference 1.