CHAPTER 3, CALIBRATION 
98307600 Rev. C-1  3-9 
1.  Connect the flowmeter to the analyzer sample inlet and measure the sample 
flow reading. Record the average of flow readings and ensure that it meets the 
specifications of section 1.1.10  ±1%.  
2.  Setup the dilution gas calibrator to obtain a CO gas concentration of 80% of 
the  EC9830’s  operating  range  as  discussed  in  section  3.2.1.4.  Perform  a 
manual calibration at this point. 
 
 
Note 
Record  the  analyzer  instruments  gains  before  and 
after the calibration. 
3.  Setup  the  dilution  gas  calibrator  to  obtain  Zero  air,  and  ensure  that  the 
EC9830 has a zero reading lower than ±5ppb NOTE: Perform a background 
otherwise and repeat step 2. 
Note 
Do  not  make  any  further  span  adjustments  during 
the  remainder  of  the  calibration.  Record  the  new 
span values in the spreadsheet. 
4.  Generate 5 additional concentrations by decreasing F
SO2
 or increasing F
D
. Be 
sure  the  total  flow  exceeds  the  analyzer’s  total  flow  demand.  For  each 
concentration generated, calculate the exact CO concentration using 3.2.1.4 
Note 
The  recommended  multipoint  span  concentrations 
are: 20, 40, 60, 80, 100% of Full Scale @ minimum 
of 1LPM.  
5.  Record the concentration and the analyzer’s response for each concentration 
after a stabilization period of 15 minutes per point. 
6.  Plot the analyzer’s responses versus the corresponding CO concentrations and 
draw or calculate the calibration curve as discussed in the following section. 
3.2.2.4 Calculating Multipoint Calibration Results 
3.2.2.4.1 Manual Calculations (Used when Excel not available) 
 
Determine the percent difference between instrument response and the calculated 
concentration using the following equation.