CHAPTER 3, CALIBRATION 
98307600 Rev. C-1  3-15 
Use  of  a  Level 1  span  calibration  (with  SPAN  COMP:  ENABLED  and 
CALIBRATION: TIMED) adjusts the instrument gain so the output of the instrument 
agrees with the concentration expected for span gas. The previously determined 
multipoint  calibration  curve is  used  to  verify that  the  analyzer output is  linear. 
Note  that  a  Level 1  span  calibration  requires  external  zero  and  span  standards 
connected via the (optional) EZS valve assembly. 
It  is  recommended  that  the  CO  gas  cylinder  be  checked  against  the  previous 
instrument  calibration  curve  immediately after  the  generation of  the  calibration 
curve (refer to Section 2.0.7 of Calibration Reference 1). It is also recommended 
that the concentration of this pollutant source be between 70% and 90% of the  
upper range limit of the analyzer and previous calibration curve. Subsequent use 
of this pollutant source, with AZS and compensation enabled, adjusts the span of 
the instrument to agree with the previous calibration line. Specific guidelines are 
contained  in  Calibration  Reference 1  for  use  of  Level 1  span  checks 
(Section 2.0.9)  and  certification  of  gas  cylinders  to  SRM/CRM  sources 
(Section 2.0.7). 
Note 
Use of SPAN COMP: ENABLED is not allowed under 
U.S. EPA designation at this time. 
A Level 2 span check (with SPAN COMP: DISABLED) does not require certification 
of the span gas used during AZS, and the result of such a check may not be used 
to correct the data, but merely serves to indicate that the analyzer is functioning 
properly. If the AZS is used for a Level 2 span check, the SPAN COMP must be set 
to  DISABLED.  A  Level 2  AZS  cycle  should  be  initiated  immediately  after 
multipoint calibration so that a valid reference point can be determined. 
As described in this section, two modes of operation are possible. In the INTERNAL 
mode,  the  internal  zero  air  scrubber  is  used  as  the  source  of  zero  air.  In  the 
EXTERNAL  mode,  the  user  must  supply  zero  air  through  the  Zero  inlet  of  the 
optional EZS valve. In the EXTERNAL  mode, the Outlet port is connected with a 
short piece of tubing to the Inlet port. 
In either mode, the user must supply the span gas. In the INTERNAL  mode, the 
span gas is  connected to the Auxiliary port of the instrument.  In the EXTERNAL 
mode, the span gas is connected to the Span port of the instrument. 
These gases (zero, if applicable, and span) must be provided to the analyzer at 
atmospheric pressure; for example, through a manifold as shown in Figure 3-1.