A386-11-880 Issue E
Page 14 © Edwards Limited 2008. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Installation
3.6 Leak-test the installation
Leak-test the system after installation and seal any leaks found. Dangerous substances which leak from the system
will be dangerous to people and there may be a danger of explosion if air leaks into the system. We recommend that
the leak rate is 1 x 10
-5
mbar.ls
-1
(1 x 10
-3
Pa.ls
-1
) helium or less.
WARNING
Leak-test the system after installation and seal any leaks found to prevent leakage of dangerous
substances out of the system and leakage of air into the system.