If you want to remove a rule, tap on it, set the Condition to “Disabled” and save the settings.
Measurement interval
Measurement interval defines how often the logger takes a measurement. It is possible to set the
measurement interval to any value from 1 minute to 10 days. Changing the measurement period does
not remove measurements from the device’s memory by default.
In order to change the measurement interval:
1. Connect to the logger, tap on the three dots in the upper right corner -> “Change measurement
period”
2. Set the new measurement interval and save changes
In order to change the measurement interval and remove all the measurements from the device’s
memory:
1. Connect to the logger, tap on the three dots in the upper right corner -> “Change measurement
period”
2. Set the new measurement interval, select “Clear memory” and save changes
Transmission interval
With this parameter you can set how often the logger sends data to Efento Cloud platform or a server.
You can set any value from 5 minutes to 10 days. The maximum value of the transmission interval
depends on the measurement interval - it can be set to 60 times the measurement interval (e.g.
measurement interval = 1 minute, transmission interval can be set to any value from 5 minutes to 60
minutes).
If the logger works with Efento Cloud, the platform informs the user, when the new data from the device
is expected.
Additionally, it is possible to configure how often the message sent by the logger requires confirmation of
receipt by the server (ACK). If the ACK is not received by the logger, it will retransmit the data to the
server. The confirmation frame (ACK) is used also to send a new configuration to the device.
Type of messages sent by the loggers depend on the use case. If your use case requires all the
messages to reach the server, then you should set the logger to always require the ACK. If your use
case is not critical, and you are aware that some data frames can be lost, you can set the device to
require the confirmation less often (e.g. every 24 hours). This will increase the battery lifetime, as the
device stays in the transmission mode for less time.