7
DO NOT continue to use if you have an allergic reaction to the device materials or if your skin appears irritated or inflamed after use. Check with
a healthcare professional before restarting use.
To reduce the risks associated with inaccurate data acquisition, store and operate this device only as instructed in this manual.
It’s recommended that the battery be recharged within 30 minutes of the low battery indicator warning. Recharge the battery using only the
appropriate USB-C cable.
DO NOT immerse the device in a liquid or subject it to any sterilization processes other than those described in this manual. The device is non-sterile.
To reduce the risk of device interference, keep the device at least two meters away from all RF emitters, including Wi-Fi routers and radios when
operating or charging.
To reduce the risks associated with very strong electromagnetic fields, avoid using the device near strong radio frequency (RF) signals or portable
and/or mobile RF devices.
If sudden or unexpected sounds are heard, move away from any radio transmitting antennas. Using accessories, transducers, and cables not
produced by Eko may result in increased RF emissions or decreased immunity.
The device contains a Bluetooth Class 2 wireless data link. The maximum radio frequency field strength generated by the device is below three
volts per meter, a level that is considered safe to use with other medical devices. However, audio, video, and other similar equipment may cause
electromagnetic interference. If such devices are encountered and cause interference, immediately move the device away from that device and/
or turn the Bluetooth feature of the interfering device OFF.
The device uses a Bluetooth Class 2 wireless data link. The Bluetooth range will be reduced when objects (walls, furniture, people, etc.) are
between the device and a paired mobile device. To improve Bluetooth connection, reduce the distance and/or allow a line of sight between the
device and the mobile device.
To reduce the risks associated with environmental contamination, follow applicable regulations when disposing of this device. The device
contains a rechargeable battery. Please properly dispose of the device as mandated by local directives.
Disperse any static electricity before using the unit.
Do not operate or store the CORE 500™ in extremely hot, cold, humid, or wet conditions.
The ECG and body sounds should be used in conjunction with a clinical evaluation. Do not use as the sole basis for medication or treatment decisions.
Never use the stethoscope without eartips firmly locked in place.
CORE 500™ is not intended for use with flammable anesthetics or flammable agents.
DO NOT use portable RF communications equipment (including peripherals such as antenna cables and external antennas) closer than 30 cm
(12 inches) to any part of the CORE 500™. Otherwise, degradation of the performance of the CORE 500 could result.
CORE 500™ is not capable of recording ECG activity of an implanted pacemaker.
CORE 500™ does not perform automated analyses or semi-automated analyses on the ECG or cardiac sounds.

Other manuals for EKO CORE 500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EKO CORE 500 and is the answer not in the manual?

EKO CORE 500 Specifications

General IconGeneral
TypeDigital Stethoscope
ConnectivityBluetooth
AmplificationUp to 40x
Active Noise CancellationYes
FDA ClearanceYes
HSA/FSA eligibilityYes
Water ResistanceIPX4
CompatibilityiOS and Android