14
Environmental Specifications
Environmental Conditions of Transport and Storage
Between Uses
– 20 °C to + 5 °C
+ 5 °C to + 35 °C, relative humidity up to 90 %, non-condensing
> 35 °C to 60 °C at a water vapor pressure up to 50 hPa
(conforming to IEC 60601-1-11 and 60601-2-47)
Continuous Operating Conditions
5 °C to + 45 °C; relative humidity range of 15 % to 90 %, non-condensing (conforming to IEC 60601-1-11 and
60601-2-47)
Ingress Protection
IP Rating 44
IP44 is protection Against ingress of solid foreign objects ≥ 1.0 mm diameter and splashing water
User Interface
Chestpiece
Hand-held device with capacitive touch, mode button, volume button, and USB-C port (charging only).
CORE 500 can be safely charged* using a Certified USB-IF, Class II Double insulated USB charging port with output
voltage rated at 4.75V-5.25V and charging current at 500mA - 2A.
*(CORE 500 was tested with an Apple Model:A1385 USB Power Adapter.)
Mobile Device iPhone with iOS 6.1 and above
Earpiece Standard 3.5 mm female TRS jack
* The heart rate algorithm has not been validated for patients under the age of 1.