EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 70
(1) Uploading Data
Assigin a number between 1 through 120 as uploading number to the measuremenet data, which
is selected from the Data List, and click the [Upload] button will upload this selected measurement
data to the MP-11 Main Unit memory. The uploaded measurement data is assigned to the
memory numbers between 301 through 420 section in the MP-11 Main Unit memory.
Assigning 0 (zero) to a measurement data will register the measurement data to a free number
above 301 from the lower numbers.
(2) Erasing Data
To delete the measurement data registered in the MP-11 Main Unit memory, enter the data
memory number and click [Erase] button. If the data memory number is 301, enter just 1,
same as the number entered when the data was uploaded.
By entering 0 (zero) and click [Erase] button, all data memory registered in the numbers between
301 through 420 are deleted all at once.
NOTE: The measurement data registered in the data memory numbers 301~420 in the MP-11
Main Unit memory cannot be deteled by the key operation on the Main Unit. Always
connect the Main Unit with PC and delete the data from the softare side.