EKO INSTRUMENTS CO., LTD. - Pyranometer MS-80/80S/80U - Instruction Manual Ver. 1
Pg. 59
A4-7. No. 0-49 Register details
The register map changes depending on the register type setting. The following table shows 4 register
maps.
See in parentheses for the unit.
Fixed value if the content is numeric.
The blank space is not used. It is filled with zero in U16 format.
The abbreviations are as follows.
Addr: Modbus slave address (1-247)
Comp Irr: Irradiance value after compensation of temperature and linearity.
Raw.Irr: Irradiance before correction
Temp: Sensor temperature
AD: Input voltage
Sensitivity: Calibration factor
Cal. Date: Calibration date
Supply: Supply Voltage
Sensor Name: Sensor name (MS-80S)
Firm Ver: Firmware version number
Hard Ver: Hardware version number
T. Coeff: Temperature compensation factor, set by No.174-181
L. Coeff: Linearity compensation factor, set by No.182-189