SERVICE MANUAL
- 6 - SM-eVENT-GB-3_8-2006.doc
3. TECHNICAL DATA
TYPE DK50 D DK50 DM
Output flow at pressure 3,5 bar
Lit.min
-1
max. 40 max. 60
Rated voltage / Frequency V/Hz 230/50
230/60*
115/60*
230/50
230/60*
115/60*
Nominal current A 2,8 2,8 5,6 3,4 4,3 8,6
Filtration of air to
µm
5 5
Pressure dew point at max.
output flow and at an air
pressure of 3 bar
°C
5°C under
the ambient temperature
5°C under
the ambient temperature
Sound level dB(A)
≤ 49 ≤ 50
Mode of the operation continuous - S1 continuous - S1
Release of condensed water automatic automatic
Indication of cooling failure acoustic and optical acoustic and optical
Output pressure bar max. 3,5 max. 3,5
Pressure tank capacity Lit 5 5
Working pressure of the pump bar 5,0– 7,0 5,0– 7,0
Operation pressure of safety
valve
bar 8,0 8,0
Adjustment of the air pressure at the
output
pressure regulator pressure regulator
Dimensions of compressor
w x d x h
mm 400x430x730 400x500x730
Dimensions of packaging
w x d x h
mm 500x530x820 500x600x820
Net weight
kg 44 55
Gross weight
kg 60 75
Implementation according EN 60601-
EN 12021
appliance of type B
class I.
appliance of type B
class I.
* The rated voltage and frequency by special order
Climatic conditions for storage and transport
Temperature –25°C to +55°C, 24 hrs +70°C
Relative air humidity 10% to 90 % (no condensation)
Relative air humidity for seaworthy packing 10% to 100% (included condensation)
Climatic conditions for operation
Temperature +15°C to +40°C
Relative air humidity up to +70%
PDF created with pdfFactory trial version www.pdffactory.com