Standard or
Method
Number
Date
Title
Notes
Voltage
Setting
ANSI/AWWA
C214-89
1990
Tape coating systems for
the exterior of steel water
pipes
Minimum voltage is 6 kV.
Use NACE RP0274
M
ANSI/AWWA
C214-89
1992
Fusion-bonded epoxy
coating for the interior and
exterior of steel water
pipes
.
V = 525 Thickness (mil)
VC, M
AS3894.1
1991
Site testing of protective
coatings. Method 1: Non-
conductive coatings -
Continuity test - High
voltage (brush) method
Testing coatings >150µm at
voltages >500 V
V =
.
250 Thickness m) / factor
VC, M
ASTM D4787
1988
Continuity verication of
liquid or sheet linings
applied to concrete
High voltage (above 900 V)
test. Set voltage below
dielectric breakdown
strength of lining. Move
probe at 0.3m/s (1 ft/s) max.
M
The voltage calculator included in the Elcometer 266 DC Holiday
Detector is programmed with the following standards:
APPENDIX A: STANDARDS
ASTM G6-83
AS3894.1:F3 1991
NACE SP0188-2006
ASTM G62-87
AS3894.1:F4 1991
NACE SP0490-2007
AS3894.1:F1 1991
ANSI/AWWA C213-91
NACE RP0274-04
AS3894.1:F2 1991
EN14430:2004
Other standards which do not derive the test voltage directly from the
coating thickness are not available within the Voltage Calculator
function. Testing to these standards is still possible however, by
setting the test voltage manually - see Section 9.2 ‘Manually Setting
the Voltage’ on page en-18.
The Elcometer 266 DC Holiday Detector can be used in accordance
with the following list of standards and test methods:
www.elcometer.com
en-35
R
Elcometer 266 Voltage Setting: VC = Voltage Calculator; M = Manual