Port 4
Reservoir Port 1
10. After 60mins remove bypass pack LC233
and insert a puri cation pack.
11. Process button (ON), normal operation
will now resume.
8. Allow to operate for 60 mins.
9. After 60 mins, press the process
button (OFF) and reconnect port 4
to reservoir port 1.
O2
Chorus 2 Operator Manual MANU40003 Version 03 3/19 Page 10
5. SANITISATION PROCEDURE
*Please be aware that during this procedure the
pre-treatment quality alarm will be active.
Note: