PURELAB Prima 7/15/30
Operator Manual

Questions and Answers

  • R
    Richard SmithAug 8, 2025
    What to do if the ELGA Water Filtration Systems output flow is below specification?
    • P
      Patricia StephensAug 9, 2025
      If the output flow of your ELGA Water Filtration System is below the specified level, you should: 1. Check the supply pressure. 2. Inspect the RO flow-rate for the unit against the graphs, which details treated water output versus temperature and feedwater pressure. 3. Ensure the inlet strainer is clean. 4. Contact a service technician to fit or replace the booster pump if one is installed.
  • D
    dawsonnicholasAug 15, 2025
    Why does the reservoir low level audible alarm sound on my ELGA Water Filtration Systems?
    • K
      kimberlyharrisonAug 15, 2025
      If the reservoir low level audible alarm sounds on your ELGA Water Filtration System: 1. Press the crossed bell button to mute the alarm. 2. The reservoir should automatically refill. 3. Check the process mimic to ensure the reservoir is filling. 4. Verify the feedwater supply and connections to the reservoir.
  • L
    Lisa TaylorAug 20, 2025
    How to troubleshoot no display message on ELGA PURELAB Prima 30 Water Filtration Systems?
    • E
      Edward BarnesAug 20, 2025
      If there is no display message on your ELGA Water Filtration System, check the following: 1. Verify the mains supply and lead. 2. Ensure the mains power is switched on. 3. Inspect the fuses in the power inlet module and replace them if they are blown.
  • J
    Jessica JenkinsAug 27, 2025
    How to reduce noise from ELGA Water Filtration Systems?
    • J
      James RuizAug 27, 2025
      To reduce noise from your ELGA Water Filtration System, open the front door and secure the pipework to stop any vibration.
  • P
    Patrick MartinezSep 1, 2025
    How to handle the pre-treatment cartridge replacement alarm in ELGA PURELAB Prima 30 Water Filtration Systems?
    • L
      Lori HinesSep 1, 2025
      When the pre-treatment cartridge replacement alarm activates on your ELGA Water Filtration System, replace the pre-treatment cartridge.

Summary

2. HEALTH AND SAFETY NOTES

2.1 Electricity

Emphasizes isolating electrical supply before maintenance and earthing the appliance.

2.2 Pressure

Details isolating water supply and releasing residual pressure before work.

4. CONTROLS

5. INSTALLATION INSTRUCTIONS

5.1 Unpacking the PURELAB Prima

Lists the items included with the PURELAB Prima unit.

5.2 Positioning the PURELAB Prima

Provides guidance on environmental conditions and mounting the unit.

5.3 Connecting up the PURELAB Prima

Details how to connect mains water, drain, and outlet tubes.

5.4 Initial Controller Set-Up

Explains the initial setup sequence for system preferences.

5.5 Initial Start Up

Guides through the steps for the initial start-up and rinsing process.

6. OPERATION

6.1 Alarm Conditions

Details the conditions under which the unit will signal alarms.

7. MAINTENANCE

Identification of Consumable

Identifies the LC140 pre-treatment cartridge and mentions the inlet strainer.

Frequency of Consumable Replacement

Recommends replacement frequencies for consumables like cartridges.

7.1 Replacing the LC140 Pre-treatment Cartridge

Provides step-by-step instructions for replacing the pre-treatment cartridge.

8. SANITIZATION PROCEDURES

9. TROUBLE SHOOTING

10. CONSUMABLES AND ACCESSORIES

11. KEY TO CONTROL PANEL

ELGA PURELAB Prima 30 Specifications

General IconGeneral
Flow Rate/Daily VolumeUp to 30 l/day
Power Supply100-240 V, 50/60 Hz
Output Water Quality18.2 MΩ·cm
Feed Water RequirementsPotable water
InstallationBenchtop
Operating Temperature Range5-35°C
TypeType I water purification system
ApplicationLaboratory use