  • M
    Mr. Michael CooperAug 16, 2025
    Why is there reduced flow from the ELGA Water Filtration Systems dispenser?
    • D
      David RussoAug 16, 2025
      If you're experiencing reduced water flow from the ELGA Water Filtration Systems dispenser, examine the inlet water strainer. Additionally, replace the POU filter or the UF filter if your system is equipped with one. If the issue persists, it may indicate a worn pump, and you should contact Customer Services.
  • K
    Kevin MurrayAug 19, 2025
    What to do if there is an alarm and flashing quality value on ELGA PURELAB Classic - US?
    • B
      Brian HahnAug 19, 2025
      If you're experiencing an alarm and flashing quality value on your ELGA Water Filtration Systems, first mute the alarm. Verify that the alarm set value is correct. Dispense approximately 10 liters of water to drain. Inspect the UF, if fitted. If the problem continues, replace the purification pack. If the issue persists, contact Customer Services.
  • B
    Brittany TaylorAug 21, 2025
    How to troubleshoot no display message on ELGA Water Filtration Systems?
    • A
      Alexis DaughertyAug 22, 2025
      If your ELGA Water Filtration Systems unit is not displaying any message, check the mains supply and lead to ensure they are properly connected. Make sure the mains power is switched on. Inspect the fuses in the power inlet module and replace them if they are blown.
  • W
    William PhillipsAug 25, 2025
    What causes the high water temperature alarm in ELGA PURELAB Classic - US?
    • B
      Belinda BrownAug 26, 2025
      If you're getting a high water temperature alarm on your ELGA Water Filtration Systems, first check that the correct alarm point is set. Then, verify that the feedwater temperature hasn't suddenly increased. Dispense some water to allow colder water to be drawn into the unit.
  • E
    Erika BeasleyAug 28, 2025
    What to do if ELGA PURELAB Classic - US Water Filtration Systems unit will not operate?
    • K
      Karen WrightAug 28, 2025
      If your ELGA Water Filtration Systems unit isn't operating, ensure that the correct purification pack is properly installed. Also, check that the purification pack has a valid consumable date and that the reminder hasn't been exceeded by 12 months.
  • J
    Jason CrossAug 31, 2025
    How to fix reservoir low level alarm on ELGA PURELAB Classic - US?
    • B
      Benjamin Contreras DDSAug 31, 2025
      If you receive a Reservoir Low Level alarm on your ELGA Water Filtration Systems, mute the alarm. Ensure the unit is fed from a reservoir and that the correct feature has been set. Check the feed to the reservoir to confirm it is operational and the reservoir is filling.
  • T
    Travis JohnsonSep 3, 2025
    How to clear purification pack change reminder alarm on ELGA PURELAB Classic - US?
    • M
      Michael RossSep 3, 2025
      If you receive a Purification Pack Change reminder alarm on your ELGA Water Filtration Systems, mute the alarm. Replace the purification pack.
  • D
    davidthomasSep 6, 2025
    How do I clear the UF change reminder alarm on ELGA PURELAB Classic - US Water Filtration Systems?
    • S
      Shane JordanSep 6, 2025
      If you encounter a UF Change reminder alarm on your ELGA Water Filtration Systems, mute the alarm. Replace the UF filter.
  • G
    Gregory LynnSep 9, 2025
    What should I do for the UV change reminder alarm on ELGA PURELAB Classic - US Water Filtration Systems?
    • J
      Julie WardSep 9, 2025
      If you see a UV Change reminder alarm on your ELGA Water Filtration Systems, mute the alarm. Replace the UV filter.
  • K
    Kevin MartinezSep 12, 2025
    How to fix a noisy ELGA PURELAB Classic - US Water Filtration Systems unit?
    • S
      Stephanie BriggsSep 12, 2025
      If your ELGA Water Filtration Systems unit is noisy, open the front door and secure the pipework to stop the vibration.

Summary

INTRODUCTION

1.1 Product Range

Lists the different PURELAB Classic models.

1.2 Use of this Manual

Explains the purpose and scope of the manual.

HEALTH AND SAFETY NOTES

2.1 Electricity

Safety precautions related to electrical power supply.

2.2 Ultraviolet Light

Safety precautions regarding UV lamp exposure.

2.4 Chemical and safety notes

General chemical handling and safety advice, including sanitization procedures.

PRODUCT AND PROCESS DESCRIPTION

3.1 Product Description

Overview of the PURELAB Classic water purification unit.

3.2 Process Description

Explains the water purification process for Classic DI.

3.3 Technical Specification

Details feedwater, dimensions, electrical, user interface, safety, and operational specs.

CONTROLS

4.1 Icons

Defines the icons used on the control panel.

INSTALLATION INSTRUCTIONS

5.1 Unpacking the PURELAB Classic

Lists the items included with the unit.

5.2 Positioning the PURELAB Classic

Guidelines for unit positioning and environment.

5.3 Connecting the PURELAB Classic

Instructions for connecting water inlet and drain tubes.

5.4 Initial Controller Set up

Comprehensive guide to initial controller setup, including menus, settings, and alarms.

5.5 Setting Up Password; Resetting Replacement Timers

Instructions for changing password and resetting consumable timers.

5.6 Initial Start Up

Steps for initial startup, including water supply, by-pass, and purification pack installation.

5.7 POU Filter Installation (LC134) - Optional Accessory

Guide for installing and bleeding the Point-Of-Use filter.

OPERATION

6.1 Intermittent Recirculation

Explains the unit's recirculation mode.

6.2 Manual Dispense

Steps to dispense water manually.

MAINTENANCE

7.1 Replacing the Purification Pack (LC186)

Step-by-step guide to replace the purification pack.

7.2 Replacing the Ultrafilter (LC169)

Instructions for removing and fitting the Ultrafilter.

7.3 Replacing the Ultraviolet Lamp (LC170)

Step-by-step instructions for replacing the UV lamp.

7.4 Replacing the Point-Of-Use Filter (LC134)

Guide for removing, replacing, and bleeding the POU filter.

SANITIZATION PROCEDURES

8.1 Tablet sanitization process

Procedure for sanitizing the unit using EfferSan™™ tablets.

8.2 Liquid sanitization

Procedure for sanitizing the unit using Minncare Cold Sterilant.

TROUBLE SHOOTING

CONSUMABLES AND ACCESSORIES

KEY TO CONTROL PANEL

11.1 Icons

Defines the icons used on the control panel.

11.2 Alarm Conditions

Lists alarm conditions and associated icons.

11.3 Replacement Timers

Details replacement timers for components.

WARRANTY;CONDITIONS OF SALE

USEFUL CONTACT DETAILS

ELGA PURELAB Classic - US Specifications

General IconGeneral
BrandELGA
ModelPURELAB Classic - US
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals