Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • J
    Joseph MartinezAug 7, 2025
    Why does Elma Other have no ultrasound but heating is OK?
    • D
      Darlene BrewerAug 7, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • M
    Michael StoneAug 12, 2025
    What to do if Elma LC20H has no ultrasound and no heating?
    • T
      Thomas BakerAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • R
    robertnortonAug 17, 2025
    What to do if Elma Other fuse is released, with no heating and no ultrasound?
    • A
      amanda60Aug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • N
    Natasha SanchezAug 22, 2025
    Why does Elma Other have no ultrasound or weak ultrasound, but heating is OK?
    • M
      Mary MorenoAug 22, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • L
    Leslie LeonardAug 27, 2025
    What to do if Elma Other has no ultrasound but heating is OK?
    • L
      Lisa JonesAug 27, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • G
    Garrett SmithSep 2, 2025
    What to do if Elma Other has no heating but ultrasound is OK?
    • A
      Alejandro JonesSep 2, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General Service Manual Information

Safety Warnings and Precautions

Crucial safety information to be read and observed before undertaking any repair work.

Troubleshooting Guide Overview

Guide to identify faults, causes, and remedies for various unit types.

Repair Procedures Overview

Step-by-step instructions for replacing faulty components for individual unit types.

Critical Safety Warnings

Details critical safety precautions for repair and maintenance procedures.

Important Operating Instructions

Cleaning Liquid Requirements

Information on filling the tank with suitable cleaning liquid and avoiding flammable liquids.

Description of Operating Elements

Trouble Shooting and Malfunctions

Troubleshooting a Faulty Unit

Describes how to check for general unit faults and safety precautions.

Diagnosing No Ultrasound;Heating Issues

Explains how to diagnose issues where the unit has no ultrasound or heating.

Ultrasound Function Faults

Details faults related to the ultrasound function and potential causes.

Heating System Faults

Explains common issues with the heating element and temperature regulation.

Trouble Shooting: T310(H) Model

Lists malfunctions, checks, possible causes, and remedies specific to the T310(H) model.

Trouble Shooting: T420(H);LC20 H Models

Lists malfunctions, checks, causes, and remedies for T420(H) and LC20H models.

Trouble Shooting: T460-T700 H;LC30-LC130 H

Malfunction troubleshooting for T460-T700H and LC30-LC130H models.

Trouble Shooting: T780-T1060 H Models

Malfunction troubleshooting for T780-T1060H models.

How to Open the Unit

Faulty Operating and Switch Elements

Replacing the Timer Component

Detailed instructions for replacing the timer component.

Replacing the Heating Switch

Step-by-step guide to replace the heating switch.

Replacing Control Thermostat (T420 H-T1060 H)

Instructions for replacing the control thermostat on specific models.

How to Replace the Heating Element

Faulty PXE Transducer Disks

Causes of PXE Transducer Disk Breakdown

Explains reasons for damage to PXE transducer disks.

Electrical Hazards with PXE Disks

Warns about electrical hazards associated with PXE disks.

How to Replace the Transducer Tank

Replacing Tank with Transducer System

Describes replacing the entire tank with a pre-mounted system.

Transducer Tank Replacement Procedure

Step-by-step instructions for replacing the transducer tank.

How to Replace the Interference Filter

How to Replace the Ultrasonic Generator

Finding the Cause of Generator Faults

Steps to identify the cause of generator malfunction.

Generator Replacement: T310-T420 H;LC20 H-LC30 H

Instructions for replacing the generator in T310-T420H and LC20H-LC30H units.

Generator Replacement: T460-T470 H;LC60 H

Instructions for replacing the generator in T460-T470H and LC60H units.

Generator Replacement: TS540-T700 H;LC130 H

Instructions for replacing the generator in TS540-T700H and LC130H units.

Generator Replacement: T780-T1060 H

Instructions for replacing the generator in T780-T1060H units.

How to Replace Generator Print Components

Generator Adjustment Procedures

Generator Adjustment: T310-T420 H;LC20 H-LC30 H

Adjustment procedure for T310-T420H and LC20H-LC30H units.

Generator Adjustment: T460-T470 H;LC60 H

Adjustment procedure for T460-T470H and LC60H units.

Generator Adjustment: TS540-T700 H;LC130 H

Adjustment procedure for TS540-T700H and LC130H units.

Generator Adjustment: T780-T1060 H

Adjustment procedure for T780-T1060H units.

Putting into Operation and Trial Run