Embrace™ Nest Operation Manual (0602-50001-002) Page 36
Periodically check the BabyWrap™ to ensure that it is free of dust, insects etc.
6.3 PERIODIC CALIBRATION CHECK
Once every three months, verify the temperature of a heated WarmPak™ with a standard
clinical thermometer to ensure that it is within the recommended operating range.
If the temperature of the heated WarmPak™ is not within the recommended operating range,
i.e. if the AccuTemp™ Heater is over- or under- heating the WarmPak™, do not use the
AccuTemp™ Heater.
Contact us using the information in the section on CUSTOMER SERVICE INFORMATION to report
the problem.
6.4 PERIODIC REPLACEMENT OF THE WARMPAK
Please replace the WarmPak™ after 6 months from the date of first use. DO NOT use the
WarmPak™ after its expiry date, under any circumstances.
NOTE
The WarmPak may turn slightly yellow over time. This is normal and will not affect product
functionality in any way.