Loading...
Blood Glucose Monitoring System
46
What What the What it What you
you see Monitor says means should do
HI “Your
blood glucose
reading is
above 600
mg/dL”
LO “Your
blood glucose
reading is
below 20
mg/dL”
“Low Battery”
“nil”
Test result is
higher than
600 mg/dL
Test result is
lower than 20
mg/dL
Batteries are
low
There are no
results stored
in the monitor
Wash and dry your hands
and repeat the test on your
ngertip with a new test
strip. If the result is still “HI,”
contact your physician or
healthcare professional
immediately.
Wash and dry your hands
and repeat the test on your
ngertip with a new test
strip. If the result is still
“LO,” contact your physician
or healthcare professional
immediately.
Change the battery
according to page 41,
Inserting (or changing) the
Battery.
Check if the date and time
on you monitor is set up.
See “Setting Up Your New
System” on page 10. Start
testing your blood glucose,
see “Testing Your Blood
Glucose” on page 29.