SAMPLE
Probe No. = 40
Meter Reading = 400 +/- 5 or 395 to 405
7.4 Insert probe in fuel to upper holes
and depress MEASURE switch. Report displayed
reading times multiplier after 3 seconds for
stabilization. (Due to the polarization of
the fuel sample the apparent reading will
continue to change. Only the reading 3
seconds after depressing the MEASURE switch
is correct).
Note: If a “1” is displayed on the left side
of the display, the conductivity level
exceeds the upper range of the meter.