100-214-294 REV. 05 49
6.5.4 Ultrasonics Operating Modes
Table 6.17 Degassing
Step Action
1 Turn Main Power switch on.
2 Press the On/Standby key to turn on the unit.
3
Default degas time is 5 minutes.
To change the degas time, press the Fn key until the Degas icon appears
and the Set Time icon flashes. Then press the Up/Down keys to change
the degas time.
4
Press Degas key once to start the degas process.
If necessary, use up/down keys to alter degas time during a degas cycle.
NOTICE
Refer to 6.5.11 Solution for information on degassing.
5
After completing the degas time, you are ready to set operating
parameters.
Table 6.18 Ultrasonic Operating Modes
Mode Action
Timed
Sonics
In Timed Sonics Mode, a timer will begin to count down and ultrasonics
will remain on until the timer reaches 0 minutes.
For instructions, see 6.5.5 Cleaning Items (Treating Samples) in Timed
Sonics Mode.
Constant
Sonics
In Constant Sonics Mode ultrasonics will remain on until the Sonics key is
pressed or power is turned off to the unit.
For instructions, see 6.5.6 Cleaning Items (Treating Samples) in
Constant Sonics Mode.
Auto
In Auto mode ultrasonics will start once set temperature is attained.
Ultrasonics will remain on until the timer reaches 0 minutes.
For instructions, see 6.5.7 Cleaning Items (Treating Samples) in Auto
Mode.