54 100-214-294 REV. 05
Figure 6.5 Draining of Units 1800 and 2800
Models 3800, 5800, and 8800 include a drain and valve kit.
Table 6.23 Draining your ultrasonic bath
Step Action
1 Place the unit to allow easy reach of the drain tube into a waste disposal unit.
2
Remove the thread protecting cap from the end of the unit’s drain pipe. This
will expose the white teflon sealing tape on the drain pipe’s threads.