2 Mounting and sensor wiring for 4-
wire remote installations
Topics covered in this chapter:
Mounting the transmitter to a DIN rail
Prepare the 4-wire cable
Wire the transmitter to the sensor
Ground the flowmeter components
2.1 Mounting the transmitter to a DIN rail
The transmitter is designed to be mounted on a 35 mm DIN rail. The DIN rail must be
grounded.
Mounting the transmitterFigure 2-1:
C
B
A
A. Spring clamp
B. DIN rail
C. Spring clamp release loop
2.1.1 Mounting multiple transmitters
If the ambient temperature is above 113 °F (45 °C) and you are mounting multiple
transmitters, mount the transmitters so they are at least 0.39 in (10 mm) apart.
Mounting and sensor wiring for 4-wire remote installations