5700 ATEX Installation Instructions
EB-20025742 Rev. AC December 2015
6 of 12
3) Parameters (continued)
3.3.1.1 for type 5700*1*E******(J3,terminals 1-2 FF) for the
connection of a FIELDBUS circuit in accordance with FISCO
model, Annex G of EN 60079-11:2012
voltage Ui DC 33 V
current Ii 380 mA
power Pi 5,32 W
effective internal inductance Li 5 μH
effective internal capacitance Ci 0,27 nF
3.3.2 for type 5700*1*E******(J3,terminals 3-4, mA)
voltage Ui DC 30 V
current Ii 484 mA
power Pi 2,05 W
effective internal inductance Li 5 μH
effective internal capacitance Ci 0,27 nF
3.3.3 for type 5700*1*E******(J3,terminals 5-6, FO/DO)
voltage Ui DC 30 V
current Ii 484 mA
power Pi 2,05 W
effective internal inductance Li 5 μH
effective internal capacitance Ci 11,27 nF
3.4 Sensor circuits for type 5700R1******** (J1 in J-box,
VDC+ RED; VDC- BLK; COM A WHT; COM B GRN); “ib”
voltage Uo DC 17,2 V
current; instantaneous Io 0,479 A
current; steady state Io 0,272 A
power Po 2,06 W
type of protection Ex ib IIC
max. external inductance Lo 154,9 H
max. external capacitance Co 333 nF
max. induct/resistance ratio Lo/Ro 17,26 H/
type of protection Ex ib IIB and IIIC
max. external inductance Lo 619,8 H
max. external capacitance Co 2,04 μF