2 Installation
2.1 Fork alignment in a pipe installation
Figure 2-1: Correct Fork Alignment for Pipe Installation
A
B
A
B
OK
OK
A. Tri Clamp process connections have a circular notch
B. Threaded process connections have a groove
2.2
Fork alignment in a vessel (tank) installation
Figure 2-2: Correct Fork Alignment for Vessel (Tank) Installation
A
B
C
OK OK
OK
A. Tri Clamp process connections have a circular notch
B. Threaded process connections have a groove
C. Flanged process connections have a circular notch