EasyManua.ls Logo

Emerson Rosemount 644R User Manual

Emerson Rosemount 644R
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Quick Start Guide
00825-0100-4728, Rev DB
April 2019
Rosemount
644H and 644R Smart
Temperature Transmitters
Device Revision 7 or Previous

Table of Contents

Other manuals for Emerson Rosemount 644R

Question and Answer IconNeed help?

Do you have a question about the Emerson Rosemount 644R and is the answer not in the manual?

Emerson Rosemount 644R Specifications

General IconGeneral
TypeTemperature Transmitter
Measurement TypeTemperature
HART ProtocolHART 7
DisplayOptional LCD
InputRTD, Thermocouple, Ohms, mV
Sensor TypeRTD, Thermocouple
Output4-20 mA HART
Power Supply10.5 to 42 VDC
Temperature Range-40 to 85 °C
Process Temperature Range-40°C to 85°C (-40°F to 185°F)
EnclosureAluminum
Enclosure RatingIP66
MountingField
Stability±0.1% of URL per year

Summary

Safety Messages

General Safety Notices

Essential safety guidelines, notices, and warnings for installation and operation.

Electrical Shock Warning

Critical safety warning regarding electrical hazards and contact with live terminals.

Configure (Bench Calibration)

Update Field Communicator Software

Procedure for updating software and connecting for bench calibration.

Verify Configuration

Traditional Interface and Fast Keys

Using the field communicator traditional interface and fast key sequences.

Calibration Constants Input

Procedure for entering Callendar Van-Dusen constants via traditional interface.

Calibration Constants via Dashboard

Procedure for entering Callendar Van-Dusen constants via device dashboard.

Set Switches

Setting Switches on 644H/644R

Instructions for setting switches on different Rosemount 644 models.

Mount the Transmitter

Transmitter Mounting Procedures

Step-by-step guide for installing the connection head and transmitter.

Install Universal Head

Procedure for mounting the transmitter with a universal head.

Rail Mount Mounting

Instructions for mounting rail mount transmitters with threaded sensors.

Wire and Apply Power

Transmitter Wiring

Diagrams and instructions for connecting sensors to the transmitter.

Powering the Transmitter

Procedures and notes for applying power to the transmitter.

Load Limitation

Information on transmitter power requirements and voltage limits.

Transmitter Grounding Options

Procedures for grounding the transmitter based on input types.

Perform a Loop Test

Performing Loop Tests

Steps to perform analog loop tests via traditional interface and device dashboard.

Product Certifications

European Directive Information

Details on EU Declaration of Conformity and its availability.

North American Certifications

USA explosionproof, intrinsic safety, and non-incendive certification details.

Canadian Certifications

Information on Canadian intrinsic safety and explosionproof certifications.

ATEX and IECEx Certifications

Details on ATEX and IECEx flameproof, intrinsic safety, and type n certifications.

Regional Certifications

Certifications for Brazil, China, EAC, and Japan.

Additional Type Approvals

Type approvals from ABS, BV, DNV, and LR.

Specification Tables

Tables detailing process temperature limits and entity parameters.