EasyManua.ls Logo

Emerson Rosemount 644R - Calibration Constants Input

Emerson Rosemount 644R
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.2 Input/verify the Callendar Van-Dusen constants on the
Field Communicator traditional user interface
If using sensor matching with this combination of a transmitter and sensor,
verify the constants input.
Procedure
1. From the Home screen, select 1 Device Setup > 3 Configuration > 2
Sensor Config > 1 Sensor 1 > 3 Cal Van-Dusen.
2. Set the control loop to manual and select OK.
3. At the Enter Sensor Type prompt, select Cal Van-Dusen
4. At the Enter Sensor Connection prompt, select the
appropriate number of wires.
5. Enter the R
0
, Alpha, Beta, and Delta values from the stainless steel
tag attached to the special-order sensor.
6. Return the control loop to automatic control and select OK.
2.3 Field Communicator device dashboard
Figure 2-2: Device Dashboard
April 2019 Quick Start Guide
Quick Start Guide 9

Table of Contents

Other manuals for Emerson Rosemount 644R

Related product manuals