Quick Start Guide
31
March 2016
Wall mount transmitter power terminals
See Figure 24 for wall mount transmitter terminal connections.
For AC powered transmitter (90
250VAC, 50/60 Hz)
Connect AC Neutral to terminal N and AC Line to terminal L1.
For DC powered transmitter
Connect DC- to terminal N and DC+ to terminal L1.
Field mount transmitter cover jam screw
For flow meters shipped with a cover jam screw, the screw should be installed
after the instrument has been wired and powered up. Follow the steps to install
the cover jam screw:
1. Verify the cover jam screw is completely threaded into the housing.
2. Install the housing cover and verify the cover is tight against the housing.
3. Using a 2.5 mm hex wrench, loosen the jam screw until it contacts the
transmitter cover.
4. Turn the jam screw an additional
1
/2 turn counterclockwise to secure the cover.
Note
Application of excessive torque may strip the threads.
5. Verify the cover cannot be removed.
7.0 Basic configuration
Once the magnetic flowmeter is installed and power has been supplied, the
transmitter must be configured through the basic setup. These parameters can
be configured through either a local operator interface or a HART communication
device. Configuration settings are saved in nonvolatile memory within the
transmitter. A table of all the parameters is located in Table 13. Descriptions of
the more advanced functions are included in the comprehensive product manual.
7.1 Basic setup
Tag
Tag is the quickest and shortest way of identifying and distinguishing between
transmitters. Transmitters can be tagged according to the requirements of your
application. The tag may be up to eight characters long.
Flow units (PV)
The flow units variable specifies the format in which the flow rate will be
displayed. Units should be selected to meet your particular metering needs.
Line size
The line size (sensor size) must be set to match the actual sensor connected to the
transmitter. The size must be specified in inches.