Emerson Process Management GmbH & Co. OHG8-16
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
8.3 Solving Problems Not Indicated by Status Messages
Situation Description Recommended Actions
Fluctuating or Invalid
Readout (continued)
Analog preamplier of af-
fected channel defective
Check measuring point
(
page 8-19)
Gas path(s) polluted
Check analysis cells and
windows for pollution
Clean polluted parts (see
service manual)
Check gas paths for pollution
and clean gas paths if need
be
Wrong pressure value used
for compensation
Set ambient pressure to pro-
per value ( page 6-60)
Sensor failure ( status
message "PressSensor“,
page 8-10)
Condensation inside gas path
Check temperature of gas
path(s)
Remove all sources of con-
densation
Keep all temperatures at
least 10 °C above dew point
Readout Damping Time
Too Long
Wrong signal damping settings
Check signal damping
(
page 6-48)
Pump rate too low
Distance between sampling
point and analyzer too long
Replace pump by external
model with higher pump rate
(operate in bypass mode,
page 4-5)
Gas path(s) polluted
Check gas path and sample
handling system for pollution
Clean gas path

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson ROSEMOUNT X-STREAM XEFD and is the answer not in the manual?

Emerson ROSEMOUNT X-STREAM XEFD Specifications

General IconGeneral
BrandEmerson
ModelROSEMOUNT X-STREAM XEFD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals