EasyManua.ls Logo

Emerson Tescom ER5000 Series - Page 152

Emerson Tescom ER5000 Series
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ER5000
152
The ERTune™ Program: Basic Features
A Typical Step Tuning Example
This section is an example of the proper sequence of steps to
follow, as well as typical settings, for a step tuning session for
the ER5000.
IMPORTANT
!
IMPORTANT
!
The PID settings downloaded during setup provide
excellent control for most applications. In some cases, minor adjustments
may be required to optimize performance. Refer to Table 1: Effect of P,
I and D Increases on Response Curve of Controller in the How It Works
section for basic tips on adjusting the PID settings.
If, however, you nd that these minor adjustments do not give
satisfactory results, zero out the controllers, refer to the Rules of Thumb
for PID Tuning as well as the detailed Tuning Tips beginning on page 149
and build your own conguration “from scratch.” You can return to the
default settings at any time by following the steps To reset the ER5000 to
its default PID settings.
Tuning is a gradual process, with repeated cycles of changing
settings and monitoring system response. Most of the steps in
this example will be repeated multiple times until the desired
response curve is seen in the Plot Screen.
The Initial Settings
Setpoint Source: ERTune™ Program
Type: Toggle
Setpoint 1: 25%
Setpoint 2: 75%
Proportional: 400
Derivative: 0
Integral: 0
Maximum: 0
Minimum: 0
Deadband: 0
To activate the Plot Screen
1. Click the Start Plot Button.
To toggle the setpoint
1. Click in the white space below the Setpoint 1 and Setpoint 2
entry elds. The words Hit Space Bar to Toggle appear.
2. Press the Space Bar. Each press toggles the setpoint.
(continued next page)

Table of Contents

Related product manuals