MESURES À PRENDRE EN CAS D'INCIDENT
Dysfonctionnement / Solutions

bouchon




Séchez et remplissez la chambre à poudre
Envoyez l’appareil à un centre de réparation agréé EMS


la pédale de commande
Attendez 1 à 2 minutes que le système décompresse

Appuyez sur la pédale de commande de votre unit dentaire

au niveau de votre unit dentaire


Dévissez le bouchon au dessus du lavabo, car de la poudre

rester sous pression
Videz la chambre à poudre, et refermez le bouchon



Appuyez sur la pédale de commande de votre unit dentaire

à main est obturée. Nettoyez la pièce à main


Envoyez l’appareil à un centre de réparation agréé EMS


poudre et du couvercle


Fuite d’eau entre le nez du raccord et la pièce à main
-
pareil, les remplacer si nécessaire
Changer de pièce à main


Contrôler la quantité de poudre dans votre appareil

Eliminer la poudre résiduelle de la pièce à main présente dans
les canaux en utilisant le EasyClean

En cas de chute de la pièce à main

Remplacer la pièce à main
Fuite d’air entre la pièce à main et l’appareil
Nettoyer la pièce à main


Fuite d’air et /ou d’eau sur le connecteur cordon

maintenance suivant les informations du fabricant.
Envoyez l’appareil complet à un centre de réparation agréé
EMS
TROUBLESHOOTING
Symptom / Solutions
Water comes into the powder chamber or leaks from the cap
Check the dental unit connection
Check the state of the o-rings of the turbine connection
Connect the unit to the dental unit (be careful, the powder
chamber must be empty)

Send the unit to an approved EMS repair center

Cut immediately the air input by releasing the foot pedal
Wait for 1-2 minutes for the system to decompress
Disconnect the unit from the dental unit
Press the foot control pedal of the dental unit
if no air comes out of the turbine connector, it is your dental
unit that causes the problem
If air comes out, it is the device that causes the problem
Unscrew the cap above the wash-hand basin, because the

can indeed remain under pressure
Empty the powder chamber and screw the cap back on
Connect the unit to the dental unit (be careful, the powder
chamber must be empty)
Connect the handpiece to the unit
Press the foot control pedal of the dental unit
If air comes through the handpiece connector, the handpiece
is blocked. Clean the handpiece
If no air comes out of the handpiece connector, the unit is
blocked
Send the unit to an approved EMS repair center
Air and/or powder is leaking from the threads of the cap
Check the seal and the cleanliness of the thread on the powder
chamber and on the cap
Replace the seal or the cap if necessary

Water leakage between the nose connection and handpiece
Check the O-rings on the nose connection of the device,
replace them if necessary
Change the handpiece


Check the amount of powder in your device
Check the air pressure of the dental unit
Remove the residual powder in the channels of the handpiece
with the EasyClean
Send the unit to an approved EMS repair center
If the handpiece falls
Check the state of the nozzle
Replace the handpiece
Air is leaking between the handpiece and the unit
Clean the handpiece
Replace the seal of the handpiece connector
Send the complete unit to an approved EMS repair center
Air and/or water is leaking from the connector of the cord
Check the connector of the cord, if it is necessary, carry out
maintenance following the information of the manufacturer
Send the complete unit to an approved EMS repair center

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS AIRFLOW HANDY 3.0 and is the answer not in the manual?

EMS AIRFLOW HANDY 3.0 Specifications

General IconGeneral
BrandEMS
ModelAIRFLOW HANDY 3.0
CategoryDental equipment
LanguageEnglish

Related product manuals