Permissible environmental conditions of use:
Temperature 10 to +35°C
Relative humidity 20 to 80%
Atmospheric pressure 700 to 1060hPa
Permissible environmental conditions of transport and storage:
Temperature -10 to +55°C
Relative humidity 10 to 90%
Atmospheric pressure 700 to 1060hPa
Operating altitude 3000m
Extended term transport and storage temperature: 0 to +40°C
Expected service life:
Main unit – 7 years
Applicator 3 years or 2,000,000 shocks whichever comes first
Essential Performance
BSEN 60601-1 defines Essential Performance as:
“Performance necessary to achieve freedom from unacceptable risk”
Functions of the EMS560, the absence or degradation of which could result
in a hazardous situation are:
Maximum frequency 22Hz
Maximum energy 180mJ (per shock)
Maximum Shocks 3000
Loss or degradation of these functions due to EM disturbances (eg.
electrostatic discharges or mains voltage dips) may cause temporary loss
of output but this is not considered to be hazardous.