Primo radial shockwave 560 – OM560EN Iss 3
31
Warnings/error messages continued –
#1001 – applicator expired
Another warning that may appear when the unit is first switched on. The
default limit for this is 1,000,000 shocks, at which time it may be advisable
to think about ordering a new inner capsule for the applicator. In fact the
applicators are capable of up to 2 million shocks so this should be taken
more as a warning than an essential action.
#801 – footswitch active on startup
This warning will appear if the switch has already been depressed (perhaps
jammed on by something lying on it) when the unit is switched on at the
mains. An image of the footswitch will appear on the screen. Removing the
obstruction to free up the switch should make this message disappear and
the device will then be ready for use. If there was no obstruction it may
indicate a fault (short-circuit) with the switch.
The following error codes are more indicative of a hardware/software
fault and will require that the user contacts a service engineer –
#1004 – inappropriate date error
Usually caused by a firmware crash or perhaps a flat clock back-up battery.
The date can be reset in the maintenance screen but requires a code to
access - contact the manufacturer.
#499 – pulses DLL unavailable
Caused by a firmware crash or hardware fault – contact the manufacturer
for service.
#500 – database error
As above – contact the manufacturer for service.