OM265EN issue 16
33
Treatment using the Monopulse applicator
Attach the Monopulse applicator to one of the electrode arms and
lock it into place by tightening the sleeve at the end of the arm
Connect one end of the BNC cable supplied to the socket on the rear
of the applicator and the other end to the BNC socket on the rear of
the unit.
Slacken the arm handwheels and position the applicator over the
treatment site so that the rim around the rim of the applicator is about
1 cm from the patient. Tighten the handwheels to prevent movement.
Make sure that Monopulse is selected for the electrode type.
To start treatment, turn the rotary control clockwise.
If the treatment time is zero, or the Monopulse applicator is not
connected, the unit will give a short alarm to indicate that the output
cannot be energised.
If the treatment time is not zero, the output of the Megapulse will be
energised and the treatment time will begin to count down. The
treatment icon will be displayed in the bottom right window of the
screen and the treatment light on the rear of the Monopulse
applicator will light. Advance the output control to the required level.
Fig. 23 Manual set-up screen

Related product manuals