Troubleshooting
Symptom Action
Unit shows “Tuning” and
“Treatment” will not begin.
Check correct electrodes
have been selected.
Check electrodes are
securely connected.
Check electrode cables are
routed as described in
relevant section.
Ensure that metal framed
beds, tables and chairs are
not in the treatment area.
Check electrodes are
positioned directly opposite
each other and either side of
the treatment area.
Unit shows “Treatment” but
no warmth can be felt by
patient and fluorescent
indicator does not light.
Check that pulse width and
frequency settings aren’t
producing a non-thermal
output (<20W average
power).
Check correct electrodes
have been selected.
Check electrodes are
securely connected.
Check power output level is
greater than 10%.
Unit shows “Over
temperature”
Leave unit switched on and
allow to cool for 15 minutes
before switching unit off and
then restarting unit.
If the actions above do not
resolve the problem please
contact the manufacturer or
an approved service agent for
further advice.